Экспортер:
Вектор-Медицинские системыto carry out cardiac electrophysiological examination;
for diagnostics and treatment of cardiac rhythm and conduction disorders.
CAPABILITIES:
Seven stimulation modes:
Jrthorhythmic electric cardiac pacing:
continuous, with fixed or linear increasing/decreasing or incrementally increasing/decreasing stimulation pulse frequency;
frequent and ultra-frequent, 300 to 1200 pulse/min;
possibility of synchronized stimulation, with calculation of electrophysiological parameters, after completion.
2. Paired electric cardiac pacing -1 to 4 extra-pulses between stimulation pulses with corresponding delays.
3. Programmed electric cardiac pacing:
formin a basic group of stimulation pulses consisting of eight pulses and 1 to 4 extra-pulses between basic groups;
programmed synchronized stimulation (stimulation start is synchronized with the R-wave of the patient`s ECG);
determination of electrophysiological parameters of the heart: sinus node function restoration time, corrected sinus node function restoration time, sinoatrial conduction time.
4. Manual electric cardiac pacing - single formation f sequence of pulses: single stimulation pulse and 1 to 4 extra- pulses.
5. Emergency stimulation - based on vital indications.
6. Relief - used for relief of arrhythmias.
7. Biologically - controlled pacing (demand mode)
POWER SUPPLY:
mains and built-in battery operation.